Publisher: School of Allied Medical Sciences, Shahid Beheshti University of Medical Sciences
  • Register
  • Login

Archives of Advances in Biosciences

  • Home
  • Journal Info
    • About the Journal
    • Publisher
    • Editorial Board
    • Indexing/Abstracting
    • Objectives and Scope
    • Privacy Statement
    • Journal History
  • Issues
    • Current
    • Archives
  • Publication Ethics
    • Code of Publishing Ethics
    • Ethical Guidelines
  • Journal Policies
    • Article Processing Charge (APC)
    • Allegations of Misconduct
    • Archiving
    • Article Withdrawal
    • Complaints Process
    • Conflict of Interest
    • Copyright & Open Access Policy
    • Editorial Policies
    • Peer Review Policy
    • Post-Publication Discussions and Corrections
    • Principles of Transparency
    • Repository
    • Using Artificial Intelligence (AI)
  • Guidelines
    • Author Guidelines
    • Reviewer Guidelines
    • Journal Designer Guidelines
    • Journal Language Editor Guidelines
  • Manuscript Templates
    • Original Article
    • Review Article
    • Case Reports
    • Short Communication
    • Policy Note
  • Submit
  • FAQ
  • Contact Us
Advanced Search
  1. Home
  2. Archives
  3. Vol. 6 No. 1 (2015): Winter
  4. Research/Original Articles

Vol. 6 No. 1 (2015)

January 2015

Determination of primary electron beam parameters in a Siemens Primus Linac using Monte Carlo simulation

  • Danial Seifi Makrani
  • Hadi Hasanzadeh
  • Tayyeb Allahverdi Pourfallah
  • Arash Ghasemi
  • Majid Jadidi
  • Hamed Babapour

Archives of Advances in Biosciences, Vol. 6 No. 1 (2015), 26 January 2015
https://doi.org/10.22037/jps.v6i1.8323 Published: 2015-02-24

  • View Article
  • Download
  • Cite
  • Statastics
  • Share

Abstract

The Monte Carlo method can be used to describe any technique that approximates solutions to quantitative problems through statistical sampling. This method is considered to be the most accurate method for dose calculation in radiotherapy. For complete modeling of a linear accelerator, it is required that the manufactured information covers all data, but some data such as primary electron energy must be indicated. The purpose of this study was to determine the best primary electron energy for 15 MV photon beam with varying the energy and FWHM. A Monte Carlo model for photon-beam output of a Siemens primus linear accelerator was validated by plotting the energy spectrum of photon beam and calculating the percentage depth dose (PDD) and beam profiles for 10×10 cm2 field. Square 10×10 cm2 field was validated by measurements in water by a farmer chamber. Linac head simulation was performed with BEAMnrc and dose calculation and 3D dose file were produced by DOSXYZnrc. The results were analyzed using MATLAB. It was found a good agreement between calculated PDD and beam profile for 15 MV photon beam using Monte Carlo simulation with primary electron energy of 11 MV and FWHM of 0.4 with maximum dose difference of 1.2% in PDD curves. In conclusion, using primary electron energy of 11 MV and FWHM of 0.4 has very good accuracy in calculating of dose distribution for 15 MV photon beam and it can be considered as a promising method for patient dose calculations.
Keywords:
  • BEAMnrc
  • Monte Carlo
  • Linac simulation
  • Beam Parameters
  • FWHM.
  • PDF

How to Cite

Seifi Makrani, D., Hasanzadeh, H., Allahverdi Pourfallah, T., Ghasemi, A., Jadidi, M., & Babapour, H. (2015). Determination of primary electron beam parameters in a Siemens Primus Linac using Monte Carlo simulation. Archives of Advances in Biosciences, 6(1). https://doi.org/10.22037/jps.v6i1.8323
  • ACM
  • ACS
  • APA
  • ABNT
  • Chicago
  • Harvard
  • IEEE
  • MLA
  • Turabian
  • Vancouver
  • Endnote/Zotero/Mendeley (RIS)
  • BibTeX
  • Abstract Viewed: 441 times
  • PDF Downloaded: 1104 times

Download Statastics

  • Linkedin
  • Twitter
  • Facebook
  • Google Plus
  • Telegram

Developed By

Open Journal Systems

Information

  • For Readers
  • For Authors
  • For Librarians
  • Home
  • Archives
  • Submissions
  • About the Journal
  • Editorial Team
  • Contact

Address: P.O. Box: 19395-4618, Darband St., Qods Sq.,Tehran, Iran.

Tel: +98-21-22707346

eISSN: 2783-1264

 Archives of Advances in Biosciences is an open access article under the terms of the Creative Commons Attribution -NonCommercial 4.0 International License.( CC BY-NC 4.0)

Powered by OJSPlus