Identification of Appropriate Probe Site in Maxillary Anterior Teeth for Electric Pulp Testing
Journal of Dental School, Shahid Beheshti University of Medical Sciences,
Vol. 39 No. 4 (2021),
13 December 2022
,
Page 119-122
https://doi.org/10.22037/jds.v39i4.38708
Abstract
Objectives This study aimed to find the appropriate probe site for electric pulp testing (EPT) in maxillary anterior teeth.
Methods Forty patients with available panoramic radiographs referred to the Endodontics Department were studied. In each patient, three intact teeth, including central incisor, lateral incisor, and canine, were selected from one quadrant of the maxillary arch. The incisal third, middle third, and cervical third of the buccal surface of the teeth were assessed. The respective teeth were isolated with rubber dam without a clamp. EPT was conducted starting with the lowest intensity, which gradually increased until the patient reported a sense of tingling, stinging, or heat. This process was repeated for all respective teeth at the three areas. Data were analyzed by the Friedman test (P<0.05), ANOVA (P<0.001), and Man-Whitney test (P>0.05).
Results The middle third of all three maxillary anterior teeth was an appropriate site for EPT. The lateral incisor had the lowest threshold response in all three areas compared with other teeth. No significant difference was observed between male and female patients (P>0.05).
Conclusion Within the limitations of this study, it may be concluded that the middle third of the buccal surface of maxillary anterior teeth is an optimal electrode placement site.
- Diagnosis
- Dental Pulp Test
- Maxilla
How to Cite
References
Lin J, Chandler NP. Electric pulp testing: a review. Int EndodJ. 2008;41(5):365-74.
Alghaithy RA, Qualtrough AJE. Pulp sensibility and vitality tests for diagnosing pulpal health in permanent teeth: a critical review. Int Endod J. 2017;50(2):135-42.
Bargale SD, Padmanabh SKD. Appropriate electrode placement site of electric pulp tester for the premolars: a clinical study. J Indian Soc Pedod Prev Dent. 2015;33(2):138-42.
Šimović M, Pavušek I, Ivanišević Malčić A, Jukić S, Prpić Mehičić G, Matijević J. Electric pulp test threshold responses in healthy incisors, canines, premolars and molars. Aust Dent J. 2018;44(1):54-9.
Peterson K, Söderström C, Kiani‐Anaraki M, Levy G. Evaluation of the ability of thermal and electrical tests to register pulp vitality. Endod Dent Traumatol. 1999;15(3):127-31.
Orchardson R, Cadden SW. An update on the physiology of the dentine–pulp complex. Dental Update. 2001;28(4):200-9.
Seltzer S, Bender I, Ziontz M. The interrelationship of pulp and periodontal disease. Oral Surg Oral Med Oral Pathol. 1963;16(12):1474-90.
Mumford JM. Thermal and electrical stimulation of teeth in the diagnosis of pulpal and periapical disease. Proc R Soc Med.1967;60(2):197-200.
Dummer P, Hicks R, Huws D. Clinical signs and symptoms in pulp disease. Int Endod J. 1980;13(1):27-35.
Cooley RL, Robison SF. Variables associated with electric pulp testing. Oral Surg Oral Med Oral Pathol. 1980;50(1):66-73.
Turedi I, Ulusoy A. Evaluation of electric pulp test thresholds and correct probe tip placement site in developing incisors: a clinical study in 1200 teeth. Eur Arch Paediatr Dent. 23(3):449-54.
Lin J, Chandler N, Purton D, Monteith B. Appropriate electrode placement site for electric pulp testing first molar teeth. J Endod. 2007;33(11):1296-8.
Udoye CI, Jafarzadeh H, Okechi UC, Aguwa EN. Appropriate electrode placement site for electric pulp testing of anterior teeth in Nigerian adults: a clinical study. J Oral Sci. 2010;52(2):287-92.
Ingle JI, Simon JH, Walton RE, Pashley DH, Bakland LK, Heithersay GS, et al. Pulpal pathology: Its etiology and prevention. Endodontics fifth ed Hamilton-London: BC Decker Inc. 2002:95-174.
Jacobson J. Probe placement during electric pulp-testing procedures. Oral Surg J. Oral Med Pathology. 1984;58(2):242-7.
Idon PI, Nwamba PN, Udoye CI, Jafarzadeh H. Does dental fluorosis affect electric pulp test response and/or optimum probe placement site?Aust Endod J. 2021;47(2):307-13.
Bender IB, Landau MA, Fonsecca S, Trowbridge HO. The optimum placement-site of the electrode in electric pulp testing of the 12 anterior teeth. J Am Dent Assoc. 1989;118(3):305-10.
Närhi M, Virtanen A, Kuhta J, Huopaniemi T. Electrical stimulation of teeth with a pulp tester in the cat. Scand J Dent Res. 1979;87(1):32-8.
- Abstract Viewed: 36 times
- PDF Downloaded: 45 times