Vagus Nerve Stimulation and External Defibrillation during Resuscitation; a Letter to Editor
Archives of Academic Emergency Medicine,
Vol. 6 No. 1 (2018),
1 Dey 2018
,
Page e27
https://doi.org/10.22037/aaem.v6i1.70
Abstract
Dear Editor;
External defibrillation in patients with implanted neuromodulatory devices is a crucial therapeutic challenge. We report a 63-year-old male patient with refractory epilepsy (RE) after recurrent ischaemic strokes in the middle cerebral artery and in the anterior cerebral artery territory 26 and 23 years ago. He received various therapeutic interventions to achieve seizure control with insufficient success. Therefore, vagus nerve stimulation (VNS) (model Pulse 102, Aspire SR, Cyberonics Inc, Houston, Texas) was applied via insertion of a pulse generator in the left upper chest in 2011. VNS stimulation settings were: output current 1.0 mA, pulse width 500 µs, frequency 30 Hz, 30s ON, 3.0 minute OFF.
A reduction of seizure frequency was achieved. Last antiepileptic therapy consisted of levetiracetam 1500 mg td, valproate 1000 mg td, and eslicarbazepine 800 mg td.
In 2016, he was admitted because of generalized seizure and aspiration pneumonia. Â During hospital stay he suffered a cardiac arrest (CA) with pulseless ventricular tachycardia (VT) caused by fulminant pulmonary artery embolism. After successful resuscitation, the patient experienced return of spontaneous circulation (ROSC). During resuscitation, biphasic electric shocks were applied using 150 Joule and subsequently 360 Joule with patches placed approximately 10 cm parasternal and at the left chest below the VNS. After successful ROSC the VNS was checked again and no malfunction could be detected. Stimulation settings were not changed. Impedance was normal. Seizures were not observed during the remaining day. Unfortunately, the patient died within one day after successful resuscitation and ROSC because of therapy refractory circulatory insufficiency.
VNS is an established therapeutic approach in treatment of TRE in children and adults to achieve reduction of seizure frequency with proven safety and efficacy. The safety of VNS in emergency situations like cardiac arrest due to VF with need of external defibrillation and application of large amounts of electrical energy is not clear. External defibrillation in VNS patients may potentially be harmful. The literature concerning external defibrillation during resuscitation or external cardioversion (EC) in patients with implanted electronic devices in neurological disorders is sparse. In patients with cardiac pacemakers external defibrillation may damage the cardiac device. EC applied to deep brain stimulation (DBS) patients may cause thalamotomy or DBS failure. Application of electroconvulsive therapy in psychiatric disorders seems to be safe. Two cases of electroconvulsive therapy (ECT) in VNS have been reported by Sharma et al. The first patient was a 66-year old female with major depression and the second one, a 57-year-old male with a history of bipolar disorder. Both had VNS for therapy refractory psychiatric illness. ECT was applied because of further worsening of the mental state without malfunction of the VNS device. To our knowledge, this is the first case of external defibrillation in a patient with VNS for TRE without alteration of the neurostimulator’s function. External defibrillation applied to VNS patients seem to be safe and effective. Several steps should be taken to minimize the electrical current flowing through the neurostimulator. First, position the patches as far away as possible from the VNS at least 10 cm each. Second, position the patches perpendicular to the VNS; use the lowest clinically appropriate output setting, and, finally, confirm that the VNS is functioning properly after defibrillation.
- emergency medicine
- epilepsy
- vagus nerve stimulation
- resuscitation
- external defibrillation
How to Cite
References
Sirven JI, Sperling M, Naritoku D, et al. Vagus nerve stimulation therapy for epilepsy in older adults. Neurology. 2000;54(5):1179-1182.
Venkatraghavan L, Chinnapa V, Peng P, Brull R. Non-cardiac implantable electrical devices: brief review and implications for anesthesiologists. Can J Anaesth. 2009;56(4):320-326.
Allen M. Pacemakers and implantable cardioverter defibrillators. Anaesthesia. 2006;61(9):883-890.
Yamamoto T, Katayama Y, Fukaya C, Kurihara J, Oshima H, Kasai M. Thalamotomy caused by cardioversion in a patient treated with deep brain stimulation. Stereotact Funct Neurosurg. 2000;74(2):73-82.
Sobstyl M, Michalowska M, Fiszer U, Zabek M. Deep brain stimulation failure due to external cardioversion in a patient with Parkinson's disease. Neurol Neurochir Pol. 2017;51(4):324-330.
Sharma A, Chaturvedi R, Sharma A, Sorrell JH. Electroconvulsive therapy in patients with vagus nerve stimulation. J ECT. 2009;25(2):141-143.
- Abstract Viewed: 267 times
- PDF Downloaded: 143 times