Comparing the Accuracy of Patient-Perceived Fever, Manual Fever Checking, and Non-Contact Frontal Infrared Thermometer; A Cross-sectional Study
学术急诊医学档案,
卷 14 编号 1 (2026),
1 十月 2025
,
第 e7 页
https://doi.org/10.22037/aaem.v14i1.2883
摘要
Introduction: Non-contact frontal infrared thermometers (NCFITs) have gained popularity following the COVID-19 pandemic, but their accuracy remains debated. This study compared the accuracy of three index tests of patient-perceived fever (PPF), manual fever checking (MFC) by emergency nurses, and NCFIT against tympanic thermometer (TT).
Methods: This single center, diagnostic accuracy study evaluated adult patients in the triage area of the emergency department (ED) between May and September 2023. Five emergency nurses (ENs) examined patients using convenience sampling with index tests, followed by TT measurement for fever detection as standard test (T≥38°C). The sensitivity, specificity, positive and negative likelihood ratios (PLR and NLR) and accuracy, of PPF, MFC, and NCFIT in detecting the fever were measured and reported with 95% confidence interval (CI). The inter-rater reliability of ENs for MFC among 10 patients (outside the study) was 100%.
Results: We included 279 patients with the mean age of 52.2 ± 16.98 years (51.97% male). TT identified 147 as febrile and 132 as non-febrile. The most sensitive index test was MFC, achieving 100% (95%CI: 97.5–100.0) sensitivity, 0 NLR, and 8.25 (95%CI: 5.7–12.1) PLR, with moderate accuracy (93.9% (95%CI: 89.9–96.0)). NCFIT demonstrated the lowest sensitivity (85.71% (95%CI: 78.9–90.9)) but the highest specificity (98.48% (95%CI: 93.6–99.5)) and accuracy (97.5% (95%CI: 93.1-99.8)). When the fever threshold for NCFIT was lowered to 37.5°C, sensitivity increased to 99% (95%CI: 97.4-99.9), NLR decreased to 0.01 (95%CI: 0-0.6), and PLR rose to 49.5 (95%CI: 20.8-90.2), while specificity remained at 98% (95%CI: 94-99.6).
Conclusion: MFC/PPF could rule out fever but required NCFIT/TT to rule it in. However, NCFIT achieved the highest NLR and PLR when its fever threshold was set at 37.5 °C.
- Fever
- Touch
- Skin temperature
- Emergency nursing
- Thermometers
##submission.howToCite##
参考
1. Malinovska A, Malinovska L, Nickel CH, Bingisser R. Prevalence and Prognosis of Fever Symptoms, Hypo-, and Hyperthermia in Unselected Emergency Patients. J Clin Med. 2022;11(1):24.
2. Broman ME, Vincent J-L, Ronco C, Hansson F, Bell M. The relationship between heart rate and body temperature in critically ill patients. Crit Care Med. 2021;49(3):e327-e31.
3. Mackowiak PA, Chervenak FA, Grünebaum A. Defining Fever. Open Forum Infect Dis. 2021;8(6).
4. Nonose Y, Sato Y, Kabayama H, Arisawa A, Onodera M, Imanaka H, et al. Accuracy of recorded body temperature of critically ill patients related to measurement site: a prospective observational study. Anaesth Intensive Care. 2012;40(5):820-4.
5. Bridges E, Thomas K. Noninvasive measurement of body temperature in critically ill patients. Crit Care Nurse. 2009;29(3):94-7.
6. Niven DJ, Gaudet JE, Laupland KB, Mrklas KJ, Roberts DJ, Stelfox HT. Accuracy of peripheral thermometers for estimating temperature: a systematic review and meta-analysis. Ann Intern Med. 2015;163(10):768-77.
7. Zhao Y, Bergmann JHM. Non-Contact Infrared Thermometers and Thermal Scanners for Human Body Temperature Monitoring: A Systematic Review. Sensors. 2023;23(17):7439.
8. Adams SD, Valentine A, Bucknall TK, Kouzani AZ. Technologies for fever screening in the time of COVID-19: a review. IEEE Sens J. 2021;22(17):16720-9.
9. Foster J, Lloyd AB, Havenith G. Non-contact infrared assessment of human body temperature: The journal Temperature toolbox. Temperature. 2021;8(4):306-19.
10. Piccinini F, Martinelli G, Carbonaro A. Reliability of Body Temperature Measurements Obtained with Contactless Infrared Point Thermometers Commonly Used during the COVID-19 Pandemic. Sensors. 2021;21(11):3794.
11. Aggarwal N, Garg M, Dwarakanathan V, Gautam N, Kumar SS, Jadon RS, et al. Diagnostic accuracy of non-contact infrared thermometers and thermal scanners: a systematic review and meta-analysis. J Travel Med. 2020;27(8):taaa193.
12. Lai F, Li X, Wang Q, Luo Y, Wang X, Huang X, et al. Reliability of Non-Contact Infrared Thermometers for Fever Screening Under COVID-19. Risk Manag Healthc Policy. 2022;15(null):447-56.
13. Masè M, Micarelli A, Falla M, Regli IB, Strapazzon G. Insight into the use of tympanic temperature during target temperature management in emergency and critical care: a scoping review. J Intensive Care. 2021;9(1):43.
14. Buckley RG, Conine M. Reliability of Subjective Fever in Triage of Adult Patients. Ann Emerg Med. 1996;27(6):693-5.
15. Li Y-W, Zhou L-S, Li X. Accuracy of tactile assessment of fever in children by caregivers: a systematic review and meta-analysis. Indian Pediatr. 2017;54(3):215-21.
16. Silva JE. The multiple contributions of thyroid hormone to heat production. J Clin Invest. 2001;108(1):35-7.
17. Jonklaas J, Bianco AC, Bauer AJ, Burman KD, Cappola AR, Celi FS, et al. Guidelines for the treatment of hypothyroidism: prepared by the american thyroid association task force on thyroid hormone replacement. Thyroid. 2014;24(12):1670-751.
18. Ross DS, Burch HB, Cooper DS, Greenlee MC, Laurberg P, Maia AL, et al. 2016 American Thyroid Association guidelines for diagnosis and management of hyperthyroidism and other causes of thyrotoxicosis. Thyroid. 2016;26(10):1343-421.
19. Kenny GP, Sigal RJ, McGinn R. Body temperature regulation in diabetes. Temperature (Austin). 2016;3(1):119-45.
20. Diagnosis and classification of diabetes mellitus. Diabetes Care. 2014;37 Suppl 1:S81-90.
21. van der Sande FM, Kooman JP, Leunissen KM. Haemodialysis and thermoregulation. Nephrol Dial Transplant. 2006;21(5):1450-1.
22. Zhukovsky DS. Fever and sweats in the patient with advanced cancer. Hematol Oncol Clin North Am. 2002;16(3):579-88.
23. Dinarello CA. THERMOREGULATION AND THE PATHOGENESIS OF FEVER. Infect Dis Clin North Am. 1996;10(2):433-49.
24. Grewal K, Krzyzanowska MK, McLeod S, Borgundvaag B, Atzema CL. Outcomes after emergency department use in patients with cancer receiving chemotherapy in Ontario, Canada: a population-based cohort study. CMAJ Open. 2020;8(3):E496-E505.
25. Axelrod L. Perioperative management of patients treated with glucocorticoids. Endocrinol Metab Clin North Am. 2003;32(2):367-83.
26. Coelho M, Luheshi G, Hopkins S, Pela I, Rothwell N. Multiple mechanisms mediate antipyretic action of glucocorticoids. Am J Physiol Regul Integr Comp Physiol. 1995;269(3):R527-R35.
27. Teller J, Ragazzi M, Simonetti G, Lava S. Accuracy of tympanic and forehead thermometers in private paediatric practice. Acta Paediatr. 2014;103(2):e80-e3.
28. EAT2012 Book of Proceedings THERMOLOGY INTERNATIONAL2012.
29. Mackowiak PA. Fever: basic mechanisms and management: Lippincott Williams & Wilkins; 1997.
30. Grünebaum A, Chervenak FA, McCullough LB, Dudenhausen JW, Bornstein E, Mackowiak PA. How fever is defined in COVID-19 publications: a disturbing lack of precision. J Perinat Med. 2021;49(3):255-61.
31. Mackowiak PA. Concepts of fever. Arch Intern Med. 1998;158(17):1870-81.
32. Mackowiak PA, Boulant JA. Fever’s upper limit. Fever: Basic Mechanisms and Management. 1997:147-63.
33. Sund‐Levander M, Forsberg C, Wahren LK. Normal oral, rectal, tympanic and axillary body temperature in adult men and women: a systematic literature review. Scand J Caring Sci. 2002;16(2):122-8.
34. Lockwood C, Conroy‐Hiller T, Page T. Vital signs. JBI Rep. 2004;2(6):207-30.
- 摘要 ##plugins.themes.ojsPlusA.frontend.article.viewed##: 84 ##plugins.themes.ojsPlusA.frontend.article.times##
- pdf (English) ##plugins.themes.ojsPlusA.frontend.article.downloaded##: 173 ##plugins.themes.ojsPlusA.frontend.article.times##