Strategy to manage T-Wave Oversensing in a Biventricular ICD
International Journal of Cardiovascular Practice,
Vol. 2 No. 2 (2017),
15 April 2017
,
Page 53-54
https://doi.org/10.21859/ijcp-020205
Abstract
T-wave oversensing is a relatively prevalent cause of intracardiac signals oversensing in patients with Implantable Cardioverter Defibrillator (ICD). Some of these oversensings are typically corrected with device reprogramming. If reprogramming fails to resolve the issue, invasive options such as repositioning the implanted lead may be necessary. We present a patient with dilated cardiomyopathy and intermittent T wave oversensing by a cardiac resynchronization therapy (CRT) that was managed by altering V-V timing.How to Cite
Akbarzadeh, M. A., Namazi, M. H., Safi, M., & Sheibani, M. (2017). Strategy to manage T-Wave Oversensing in a Biventricular ICD. International Journal of Cardiovascular Practice, 2(2), 53–54. https://doi.org/10.21859/ijcp-020205
- Abstract Viewed: 423 times
- PDF Downloaded: 518 times