Evaluating the Repeatability and Effect of Sex and Age on Corneal Biomechanical Parameters Measured Using Corvis ST
Journal of Ophthalmic and Optometric Sciences,
Vol. 2 No. 1 (2018),
1 Dey 2018
,
Page 14-21
https://doi.org/10.22037/joos.v2i1.28560
Abstract
Abstract
Purpose: To evaluate the repeatability of biomechanical readings by Corvis ST (Wetzlar, Germany) in healthy eyes and its relation with age and sex.
Methods: Three consecutive measurements were performed on 100 eyes of 100 patients using the Corvis ST. Various parameters including first and second applanation length, first and second applanation velocity, first and second applanation time, peak distance, radius, deformation amplitude, corrected and non-corrected intraocular pressure (IOP), central corneal thickness (CCT), Corvis ST biomechanical index (CBI), and tomography and biomechanical index (TBI) were derived from the Corvis ST readings. Repeatability of each parameter and their correlation with age and sex were evaluated.
Results: The mean IOP, CCT, CBI and TBI were 15.47 ± 2.24 mm Hg, 541.24 ± 38.90 μm, 0.048 ± 0.13 and 0.138 ± 0.17 respectively. Intra class correlation coefficient (ICC) for CCT, second applanation velocity and deformation amplitude were 0.985, 0.809 and 0.825 respectively. ICC was less than 0.8 for all other parameters. The device-specific readings showed no significant relationship with age and sex.
Conclusion: The Corvis ST showed high repeatability for CCT, second applanation velocity, and deformation amplitude parameters. No relation between Corvis ST readings and age or sex was observed.
Keywords: Cornea; Biomechanical; Age; Sex; Iran.
How to Cite
References
Garcia-Porta N, Fernandes P, Queiros A, Salgado-Borges J, Parafita-Mato M, González-Méijome JM. Corneal biomechanical properties in different ocular conditions and new measurement techniques. ISRN Ophthalmol. 2014;2014:724546.
Kling S, Hafezi F. Corneal biomechanics - a review. Ophthalmic Physiol Opt. 2017;37(3):240-52.
Luce DA. Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg. 2005;31(1):156-62.
Medeiros FA, Weinreb RN. Evaluation of the influence of corneal biomechanical properties on intraocular pressure measurements using the ocular response analyzer. J Glaucoma. 2006;15(5):364-70.
Hon Y, Lam AK. Corneal deformation measurement using
Scheimpflug noncontact tonometry. Optom Vis Sci. 2013;90(1):e1-8.
Huang J, Ding X, Savini G, Pan C, Feng Y, Cheng D, et al. A Comparison between Scheimpflug imaging and optical coherence tomography in measuring corneal thickness. Ophthalmology. 2013;120(10):1951-8.
Dharwadkar S, Nayak B. Corneal topography and tomography. Journal of Clinical Ophthalmology and Research. 2015;3(1):45-62.
Tian L, Huang YF, Wang LQ, Bai H, Wang Q, Jiang JJ, et al. Corneal biomechanical assessment using corneal visualization scheimpflug technology in keratoconic and normal eyes. J Ophthalmol. 2014;2014:147516.
Ambrósio R Jr, Lopes BT, Faria-Correia F, Salomão MQ, Bühren J, Roberts CJ, et al. Integration of Scheimpflug-Based Corneal Tomography and Biomechanical Assessments for Enhancing Ectasia Detection. J Refract Surg. 2017;33(7):434-43.
Vinciguerra R, Ambrósio R Jr, Elsheikh A, Roberts CJ, Lopes B, Morenghi E, et al. Detection of Keratoconus With a New Biomechanical Index. J Refract Surg. 2016;32(12):803-10.
Luz A, Faria-Correia F, Salomão MQ, Lopes BT, Ambrósio R Jr. Corneal biomechanics: Where are we? J Curr Ophthalmol. 2016;28(3):97-8.
Moshirfar M, Edmonds JN, Behunin NL, Christiansen SM. Corneal biomechanics in iatrogenic ectasia and keratoconus: A review of the literature. Oman J Ophthalmol. 2013;6(1):12-7.
Ali NQ, Patel DV, McGhee CN. Biomechanical responses of healthy and keratoconic corneas measured using a noncontact scheimpflug-based tonometer. Invest Ophthalmol Vis Sci. 2014;55(6):3651-9.
Wang W, He M, He H, Zhang C, Jin H, Zhong X. Corneal biomechanical metrics of healthy Chinese adults using Corvis ST. Cont Lens Anterior Eye. 2017;40(2):97-103.
Nemeth G, Hassan Z, Csutak A, Szalai E, Berta A, Modis L Jr. Repeatability of ocular biomechanical data measurements with a Scheimpflug-based noncontact device on normal corneas. J Refract Surg. 2013;29(8):558-63.
Strobbe E, Cellini M, Barbaresi U, Campos EC. Influence of age and gender on corneal biomechanical properties in a healthy Italian population. Cornea. 2014;33(9):968-72.
Allam RS, Khalil NM. Evaluation of sex differences in corneal hysteresis. Eur J Ophthalmol. 2015;25(5):391-5.
Valbon BF, Ambrósio R Jr, Fontes BM, Luz A, Roberts CJ, Alves MR. Ocular biomechanical metrics by CorVis ST in healthy Brazilian patients. J Refract Surg. 2014;30(7):468-73.
Asaoka R, Nakakura S, Tabuchi H, Murata H, Nakao Y, Ihara N, et al. The Relationship between Corvis ST Tonometry Measured Corneal Parameters and Intraocular Pressure, Corneal Thickness and Corneal Curvature. PLoS One. 2015;10(10):e0140385.
- Abstract Viewed: 213 times
- PDF Downloaded: 129 times